Move a slider to watch the Nyquist and both Bode plots update together. The solid teal curve is the one you change. The dashed purple curve is your reference. Use the baseline to see cumulative changes.
Before you explore · measurement setup
Choose a measurement configuration that matches the question. In a two-electrode setup, the measured impedance includes both electrode interfaces and solution paths. In a three-electrode setup, the working electrode is measured against a nearby reference electrode while a counter electrode carries current; placement and cell geometry still affect the result.
Set and record
Reference and counter electrode identity, placement and geometry.
DC bias or open-circuit potential, and AC amplitude.
Frequency limits, points per decade, stabilization and repeats.
Electrode area, electrolyte, temperature and instrument.
Validate the response
Use an amplitude small enough for the local linear regime; confirm by checking amplitude dependence when needed.
Check stationarity with repeat or reverse sweeps, especially at low frequency.
Inspect leads, contacts, parasitic inductance, range limits, noise and sign convention.
Compare Nyquist and Bode plots before fitting.
For the experiment-to-publication checklist and reporting fields, open .
1 · Select a circuit
2 · Adjust parameters
3 · Observe the impedance spectrum
Nyquist · −Im(Z) vs Re(Z)
Bode · Impedance magnitude |Z|
Bode · Phase angle
Current circuitSaved baselineFrequency: 0.1 Hz–100 kHz · logarithmic spacingAxes adapt dynamically during slider movement
We use standard electrochemical names consistently. Each concept includes a plain-language explanation, a picture of the element or process, an example of the expected graph, a formal definition, and a limitation. Graphs are illustrative idealized examples, not measured spectra.
Start here · What EIS measures
Electrochemical impedance spectroscopy applies a small alternating voltage and measures the resulting current across a range of frequencies. The ratio between voltage and current is impedance, Z. A resistor has the same response at every frequency; capacitive, inductive and distributed processes change their response with frequency.
From a sine wave to impedance
For a sinusoidal perturbation, voltage and current can differ in amplitude and timing. Impedance combines both: Z = Ṽ / Ĩ = Z′ + jZ″. Its magnitude |Z| describes the amplitude ratio; phase φ describes the lead or lag. For a resistor, Z = R and φ = 0°. For an ideal capacitor, Z = 1/(jωC), |Z| = 1/(ωC), and φ = −90°, where ω = 2πf.
The smaller the perturbation, the more likely the measurement represents local linear behavior around the chosen DC bias or open-circuit potential. Linearity and a stable sample during the sweep must still be checked.
Two views of the same data
Nyquist: plots −Z″ against Z′ and reveals loci such as ideal RC semicircles. Bode: plots |Z| and phase against frequency on a logarithmic axis. The same process can look different in each view, so use them together.
Frequency sets the timescale being probed: high frequencies emphasize fast response and series contributions; lower frequencies reveal slower relaxations, diffusion-like behavior or drift. These are clues, not unique physical assignments.
Read the symbol, then ask what evidence supports its name. An R is mathematically an R whether named Rₛ, Rₚ or Rct. Naming it does not establish a mechanism. Select any family for definitions, equations, limits and a graphical example.
Spatially distributed resistance and capacitance, not a single element.
Three levels of inference: observed spectral shape → mathematical model that reproduces it → physical mechanism supported by additional experiments. A fitted label alone establishes only the middle step.
How to interpret resistance names: Rs, Re, Rp, Rct and Rfilm
What does polarization resistance mean?
Polarization is a change in electrode potential away from a chosen operating or equilibrium condition when current flows. Polarization resistance is a local differential slope of the potential–current relationship, after stating how uncompensated ohmic effects are handled:
Rp = dηdi
Here η is overpotential and i is current. If current density is used instead, the area-normalized resistance has units Ω·cm² (or another area unit). In a simple charge-transfer-controlled reaction near equilibrium, Rp may coincide with Rct under the model assumptions. In more complex systems it can include several kinetic contributions.
Notation caution: an author may also write Rp for the resistor in a parallel branch without establishing a measured polarization resistance. Read the circuit definition before interpreting the symbol.
Rct: resistance associated with interfacial electron-transfer kinetics in a specified reaction model. Without evidence for the reaction, use a neutral branch label such as R1 rather than assuming Rct.
Rs, Re, RΩ: often electrolyte/series/ohmic contributions, but their exact scope varies by author. Ru: uncompensated ohmic resistance between the working interface and reference sensing point. Rfilm: a proposed film-associated resistance that requires a justified film model.
When different elements give the same spectrum
Change the CPE exponent to compare its frequency response with an ideal capacitor and a semi-infinite Warburg element. All curves are normalized to the same impedance magnitude at 1 Hz so their shapes can be compared.
Diffusion signatures: semi-infinite versus finite-length response
At sufficiently high frequency, diffusion models can produce a 45° Nyquist tail. A CPE with n = 0.5 has the same frequency exponent as semi-infinite Warburg impedance, so that slope alone cannot establish diffusion. A finite diffusion domain changes the low-frequency response.
Models: Z∞ = Rd/√(jωτ); ZT = Rd tanh(√(jωτ))/√(jωτ); ZO = Rd coth(√(jωτ))/√(jωτ). Here Rd = 1 kΩ. The semi-infinite and finite models share their high-frequency asymptote.
━━ Semi-infinite ━━ Selected finite boundary
Interpretation caution: These are idealized diffusion boundary models, not proof that an observed tail comes from diffusion. Roughness, distributed time constants, electrode porosity, and overlapping processes can also affect the shape. Finite-space and finite-length terminology varies; the defining boundary condition and equation are more reliable than the label.
When two relaxation processes overlap
Two ideal parallel RC branches in series can look like one broad arc when their characteristic frequencies are close. Move their separation to see when the two processes become resolvable. This is a synthetic demonstration, not a method for counting physical layers.
Circuit: Rₛ + (R₁ ∥ C₁) + (R₂ ∥ C₂). Rₛ = 100 Ω; R₁ = 1 kΩ. Center frequency f₁ = 100 Hz; f₂ varies with separation. Each branch contributes a relaxation, but its arc may be hidden in the sum.
Interpretation: A shoulder or broadened peak may suggest overlapping timescales, but a single CPE can sometimes describe similar data. Compare residuals, parameter uncertainty, repeatability and wider reliable frequency measurements before preferring a more complex circuit. Two fitted branches do not establish two physical coatings.
From an ideal arc to a depressed arc
Compare Rₛ + (R₁ ∥ C₁) with Rₛ + (R₁ ∥ CPE₁). The same resistor can appear in both models; changing the CPE exponent changes the arc shape without establishing a particular physical mechanism.
Fixed: Rₛ = 100 Ω; R₁ = 2 kΩ; Q = 10 µS·sⁿ (numerical value). The comparison uses the same numerical Q at each exponent, not a fixed capacitance.
━━ Ideal capacitor (n = 1) ┄┄ CPE (selected n)
Read the plot: the high-frequency intercept approaches Rₛ, the low-frequency intercept approaches Rₛ + R₁, and the depressed arc is lower than the ideal semicircle. These limits require a sufficiently broad frequency window. For a CPE branch, the frequency of maximum −Im(Z) is 12π(R₁Q)1/n, not the ideal RC formula unless n = 1.
Interpretation: n below 1 describes a distributed/non-ideal response. It does not, by itself, prove roughness, porosity, diffusion, or charge-transfer kinetics. Q has units S·sⁿ, so it is not generally a capacitance in farads.
Read a semicircle: what does each fitted parameter change?
Ideal teaching circuit: Rs + (R1 ∥ C1). The branch labels are neutral: an arc alone does not justify calling R1 charge-transfer resistance or C1 double-layer capacitance.
Z(ω) = Rs + R11 + jωR1C1
Arc diameter = R1; left intercept = Rs; right intercept = Rs + R1. The arc apex occurs at fpeak = 1/(2πR1C1) for this ideal model.
These exact intercept and apex relationships require a resolved ideal R∥C arc with the assumed series topology. CPEs, overlapping arcs, and limited frequency windows can change what is identifiable.
CPEIdeal capacitorSemi-infinite Warburg
Bode phase
Nyquist (normalized)
These are isolated ideal elements. Real electrode spectra include series resistance and often overlapping processes. Mathematical equivalence does not identify the physical mechanism.
Full notation index, searchable definitions and example graphs
Element notation dictionary
Names describe a circuit position or a proposed mechanism. Select a term to read its definition, what would support the interpretation, and what cannot be concluded from a fit.
Symbol
Meaning
Definition and interpretation
Equivalent mathematical behavior does not establish an equivalent mechanism. A CPE with n = 1 is an ideal capacitor (C = Q); n = 0 is a resistor (R = 1/Q); n = 0.5 has the frequency dependence of a semi-infinite Warburg element. These are mathematical limits, not automatic evidence of double-layer charging or diffusion.
What a fitted element does — and does not — identify
What a fit tells you
What the name implies
What else you need
A resistor R₁ fits an arc
Rct: resistance associated with electron-transfer kinetics
Evidence for an applicable Faradaic reaction and a suitable kinetic model
A capacitance-like branch fits
Cdl: interfacial double-layer capacitance
Evidence that this is the dominant origin of the fitted capacitance
A 45° Nyquist region fits
Warburg: diffusion-controlled response under defined assumptions
Transport mechanism, boundary conditions, and frequency-window evidence
Two time constants fit
Two physical layers or interfaces
Independent structural or experimental evidence; two branches do not prove two layers
Glossary cards and illustrative Nyquist examples
Explore related concepts
Physical properties derived from EIS models
Separate three kinds of quantity: directly fitted parameters (for example R, C, Q and n within a chosen circuit), derived properties calculated from fitted values plus geometry or other assumptions, and independent measurements needed to identify material properties. A fitted resistance alone does not determine conductivity.
Reading each entry: equation · variable definitions and SI units · physical meaning · assumptions · worked example · limitation.
Resistance and conductance
G = 1R
Variables: R resistance (Ω); G conductance (S = Ω⁻¹). Meaning: opposition to current versus ease of current flow.
Assumptions: linear response at the stated operating point; an identifiable circuit contribution. Example: R = 2.0 kΩ gives G = 0.500 mS.
Limit: a fitted branch R is topology-dependent and does not alone identify a material or mechanism.
Resistivity and conductivity
R = ρ ℓA; σ = 1ρ
Variables: ρ resistivity (Ω·m); ℓ current-path length (m); A cross-section normal to current (m²); σ conductivity (S·m⁻¹).
Meaning: bulk material transport properties in a uniform conductor. Assumptions: known geometry, homogeneous isotropic material, uniform current density, ohmic response, and R assigned to that material path.
Example: if a uniform film has R = 100 Ω, ℓ = 1.0 mm and A = 1.0 mm², then ρ = RA/ℓ = 0.10 Ω·m and σ = 10 S·m⁻¹.
Limit: electrode interface, electrolyte, contact, pore, and spreading resistances violate this simple geometry unless separately resolved.
Capacitance and areal capacitance
C = dQedV; CA = CA
Variables: Qe stored charge (C); V potential (V); C capacitance (F); CA area-normalized capacitance (F·m⁻²); A declared area (m²).
Meaning: incremental charge response to potential. Assumptions: specified potential range and frequency, meaningful area, and a model where a lumped capacitance is appropriate.
Example: C = 20 µF over geometric A = 1.0 mm² gives CA = 20 F·m⁻² (2.0 mF·cm⁻²).
Limit: geometric, electrochemically active, and roughness-corrected areas differ. State which one is used. A CPE Q is not in general C.
Mass-specific capacitance
C is capacitance in farads, m is the independently measured active mass in kilograms, and Cₘ is capacitance per mass in F·kg⁻¹. Use a capacitance attributable to that material at the stated bias and measurement timescale.
Example: 0.50 F from 2.0 mg of active material gives 250 F·g⁻¹. State the mass convention and whether the reported capacitance describes one electrode or the whole cell. A fitted CPE coefficient cannot be normalized as capacitance without a justified conversion.
Permittivity
C = ε Ad; εr = εε0
Variables: ε permittivity (F·m⁻¹); εr relative permittivity (dimensionless); ε0 vacuum permittivity; A plate area (m²); d separation (m).
Meaning: dielectric polarization in a simple field geometry. Assumptions: parallel plates, uniform dielectric, negligible fringing and leakage.
Example: C = 8.85 pF, d = 1.0 mm, A = 1.0 cm² gives ε ≈ 8.85×10⁻¹¹ F·m⁻¹ and εr ≈ 10.0.
Limit: an interfacial electrode capacitance usually does not satisfy this geometry; independently known thickness and field model are needed.
RC relaxation time and characteristic frequency
τ = RC; f0 = 12πτ
Variables: τ time constant (s); R (Ω); C (F); f0 characteristic frequency (Hz).
Meaning: for an ideal parallel RC branch, f0 is the −45° phase point of the branch impedance and the +45° phase point of its admittance and the Nyquist arc apex. A characteristic frequency marks a model timescale.
Assumptions: single ideal RC relaxation, stated topology, and the observed feature belongs to that branch. Example: R = 2.0 kΩ, C = 10 µF gives τ = 20 ms and f0 = 7.96 Hz.
Limit: −45° is not a universal cutoff definition for a total multi-process spectrum. Multiple crossings can result from combined dispersions, series terms or overlapping processes; crossing count is not process count.
Effective capacitance from an R ∥ CPE branch
Ceff = Q1/n R(1−n)/n
Variables: Q CPE coefficient (S·sn); n dimensionless exponent; R parallel resistance (Ω); Ceff effective capacitance (F).
Meaning: one common conversion for an R ∥ CPE circuit. For this convention, ωmax = (RQ)−1/n and Ceff = 1/(Rωmax).
Assumptions: the chosen parallel topology, 0 < n ≤ 1, Q units consistent with n, and a resolved relaxation. Example: Q = 10 µS·s0.8, n = 0.8, R = 2.0 kΩ gives Ceff ≈ 3.76 µF.
Limit: conversions depend on circuit topology and convention; Ceff is model-derived and is not a unique microscopic capacitance.
Charge-transfer resistance and kinetics
Rct = RTneF i0 (small-signal Butler–Volmer)
Variables: R gas constant (J·mol⁻¹·K⁻¹); T (K); ne electrons per reaction; F Faraday constant (C·mol⁻¹); i0 exchange current (A).
Meaning: differential kinetic resistance near equilibrium for a specified reaction and electrode area convention. Area-specific Rct (Ω·m²) relates to exchange-current density j₀ (A·m⁻²).
Assumptions: small overpotential, equilibrium, Butler–Volmer kinetics, identified reaction, and consistent total-current or current-density basis.
Example: at 298 K, ne=1 and i₀=1 mA, Rct ≈ 25.7 Ω. Limit: do not call a generic fitted arc resistor Rct without reaction-specific support.
Electrode-specific quantities
ZA = ZA; CSC = QcA; CIC = |Qphase|A
Variables: Z impedance (Ω); A stated electrode area (m² or cm²); ZA area-normalized impedance (Ω·m²); Qc cathodic charge (C); CSC cathodal charge storage capacity (C·m⁻²).
Meaning: geometry-normalized comparison metrics, only comparable when area convention, electrolyte, potential, frequency and protocol are matched.
Assumptions: report geometric versus true/roughness area, exposed area, and any coating boundary. Example: 1 kΩ·mm² for 1 mm² gives 1 kΩ·mm² area product; 2 µC over 1 mm² gives 2 C·m⁻².
Additional metrics: report |Z| at 1 kHz as a measured comparison point with matched conditions; define any application-specific cutoff explicitly from the relevant transfer function. Charge-injection capacity (CIC) is often expressed as Qphase/A (C·m⁻²) for a stated pulse waveform and potential limits.
Example: Qphase = 2 µC over 1 mm² gives CIC = 2 C·m⁻² for that specific pulse protocol.
Limit: area normalization does not remove geometry, access resistance, edge effects or material differences. CSC is protocol-dependent and is not interchangeable with a small-signal EIS capacitance; CIC requires pulse testing and a declared safe potential window.
Which quantities require independent measurements?
Equivalent-circuit fitting can directly return the parameters defined by the selected model. Derived quantities require an explicit equation plus model assumptions. Conductivity/resistivity require a known current path and geometry; permittivity requires a defensible dielectric field geometry; Rct kinetics require an identified Faradaic reaction and kinetic model; specific capacitance requires a declared area convention. EIS alone often cannot supply these independently.
Two processes can occur together: non-Faradaic charging rearranges charge near the surface, while Faradaic reactions transfer electrons in an electrochemical reaction. These are mechanisms, not interchangeable circuit labels. [1] EIS tutorial ↗
Non-Faradaic · charging without an electron-transfer reaction
Analogy: Two groups of people stand on opposite sides of a fence. They can move closer or farther apart without crossing the fence.
Physical picture: Electronic charge in the metal and ionic charge in solution rearrange. This creates double-layer charging current; it does not require an interfacial electron-transfer reaction.
EIS clue: A capacitive phase and frequency-dependent |Z|. A real interface may resemble a CPE rather than an ideal capacitor.
Faradaic · electrochemical electron transfer
Analogy: This time a person actually passes an object through the fence, changing what is held on each side.
Physical picture: An oxidized species accepts an electron (reduction), or a reduced species releases an electron (oxidation). Reaction kinetics and reactant transport can affect impedance.
EIS clue: A resistive arc and/or diffusion-like feature may occur, but neither proves a redox reaction.
Why this matters for PBS, Au and PEDOT
Au/Pt in PBS without an added redox probe: Interfacial charging is often important. Faradaic reactions are still possible depending on potential, dissolved species and surface chemistry; a fitted parallel resistance must not automatically be called Rct.
PEDOT: Ions can move into and out of the polymer while its electronic oxidation state changes. This volumetric mixed ionic–electronic charging is not necessarily captured by a single double-layer capacitor or a simple charge-transfer resistor.
Evidence ladder: (1) observe a spectral feature → (2) fit a descriptive circuit → (3) test a mechanistic hypothesis using potential dependence, electrolyte changes, CV, controls and reproducibility.
Use this sequence before assigning a physical circuit. Each observation is a clue, not a unique mechanism.
1 · High-frequency limit
Check whether the real impedance approaches a resolved plateau. For an ideal series resistance, Re(Z) → Rs. Do not equate the highest measured point with Rs when no plateau is visible.
2 · Dispersion and time scale
Inspect both |Z| and phase over the full measured range. A phase feature can indicate a relaxation, but overlapping processes may produce one broad feature.
3 · Low-frequency response
Look for a second resistive limit, capacitive trend, diffusion-like tail, or drift. Low-frequency data are especially vulnerable to nonstationarity.
4 · Candidate model and residuals
Fit the simplest defensible candidate, examine complex residuals, test alternatives, and report uncertainty. A good numerical fit alone does not establish the mechanism.
Reference equations — idealized limiting cases
ZR = R
ZC = 1jωC
ZCPE = 1Q(jω)n
ZR∥C = R1 + jωRC
Here ω = 2πf. For the ideal R∥C branch, the characteristic frequency is 12πRC. These formulas do not identify a physical process without supporting evidence.
What does 1 kHz miss?
Two electrodes can have the same |Z| at 1 kHz yet differ at 10 Hz. The point is not that every electrode needs a −45° cutoff. The point is that a full frequency spectrum reveals behavior a single-number comparison misses. A phase crossing is only a descriptive landmark unless a specific model or recording transfer function gives it a defensible meaning.
Interpretation rule: If phase crosses −45° more than once, list the crossings and inspect the corresponding spectral regions; do not automatically designate the first or last as the cutoff.
Characteristic frequency is a model timescale
A characteristic frequency is a frequency associated with a defined transition or extremum in a stated model. For an ideal parallel RC branch, f0 = 1/(2πRC) identifies the arc apex and −45° branch phase. Other models define their characteristic frequency differently. It is not automatically a device cutoff.
A −45° crossing of the total measured phase is not a universal cutoff-frequency definition. Several crossings can occur when resistive offsets and two or more dispersions combine; a crossing can shift or disappear when the measured frequency window changes. Report the chosen operational definition and its circuit/transfer-function context. Do not treat crossing count as physical process count.
For neural recording or stimulation, a cutoff must be defined from the relevant amplifier, electrode–tissue system, or application transfer function and validated in that context. A single EIS phase crossing alone does not establish that bandwidth.
EIS frequencies probe timescales, not isolated physical components. These ranges are teaching landmarks, not universal boundaries. The same process can appear at different frequencies depending on electrode size, materials, electrolyte and potential.
Explore the measured window: 0.1 Hz–100 kHz
10⁻¹ Hz10¹ Hz10³ Hz10⁵ Hz
Period: 1 ms
Frequency landmarks (illustrative, not universal)
The last column flags interpretations that the measurement does not establish on its own. These frequency windows overlap.
Frequency
What to examine
Common misunderstanding to avoid
10–100 kHz
High-frequency resistive intercept; solution/access resistance if a plateau is resolved; wiring artifacts
The highest-frequency data point is always Rs; check for a resolved intercept and artifacts
1 kHz is the frequency of an action potential, or that |Z| at 1 kHz alone predicts recording quality
10–1000 Hz
Interfacial charging, RC/CPE dispersions and overlapping relaxations
Every phase feature proves a separate chemical reaction
0.1–10 Hz
Slow charging, transport, leakage or Faradaic kinetics when supported
Every 45° tail proves molecular diffusion
A process with characteristic time τ often becomes prominent around f ≈ 1/(2πτ) for a simple RC relaxation; real porous/PEDOT electrodes can have broad distributions instead.
Foundations: [1] ↗ · [2] ↗. The 1 kHz neural-electrode benchmark is a convention, not a unique physiological frequency.
Continue with frequency-dependent interpretation
Interactive lab · which frequencies identify R and C?
For an ideal Rₛ + (R₁ ∥ C₁) branch, changing R₁ and C₁ does not affect every frequency equally. This exercise computes the normalized complex sensitivity of impedance to each parameter and shows how much of that sensitivity your measurement window captures. Sensitivity is not the same as statistical identifiability: parameter correlation, noise and model mismatch still matter.
What is being plotted? Equations and limitations
For ω = 2πf, the parallel branch has impedance Zₚ = R₁ / (1 + jωR₁C₁). The plotted sensitivities are the magnitudes of R₁ ∂Z/∂R₁ and C₁ ∂Z/∂C₁, each divided by R₁ so that the two curves share a dimensionless scale. The series resistance has no effect on these derivatives, but it changes relative measurement noise and the ability to estimate the full model.
The sensitivity to R₁ is 1 / [1 + (ωR₁C₁)²]; the sensitivity to C₁ is ωR₁C₁ / [1 + (ωR₁C₁)²]. These expressions describe the magnitude of complex derivatives, not separate real and imaginary sensitivities or a Fisher-information analysis. A good frequency window is necessary but not sufficient for reliable parameter estimation.
Use these as illustrative simulations, not experimental observations. Each example links to the live circuit playground so you can change one parameter and see its effect.
Example 1 · A mostly polarizable bare metal electrode
Observation: Predominantly capacitive response, without a clearly resolved semicircle. First candidate: Rs + CPE. Why: It represents a resistive electrolyte path and a non-ideal interfacial charging response.
Check: Does the fitted exponent n approach 1? Are low-frequency drift and high-frequency lead artifacts absent? The CPE is a descriptive element, not proof of a particular surface morphology.
Example 2 · A resolved interfacial semicircle
Observation: A single semicircle with high- and low-frequency intercepts. First candidate: Rs + (Rp ∥ Cdl). This is mathematically identical to a simple Randles circuit when Rp can be justified as charge-transfer resistance Rct.
Check: Does an independently supported redox reaction exist? Without one, keep the generic name Rp. Change Rp to see the arc diameter; change Cdl to see its characteristic frequency shift.
Example 3 · A diffusion-like low-frequency tail
Observation: A low-frequency response resembling a 45° Nyquist line. Candidate: A Warburg-containing model, only if diffusion is supported by chemistry and independent tests.
Check: Distributed charging in rough or porous electrodes can look similar. PEDOT/hydrogel systems may require mixed ionic–electronic transport models rather than a single Warburg element.
Choose the best initial descriptive model for a synthetic spectrum. These cases are deliberately idealized: real spectra can be non-unique, and a model fit does not establish the underlying chemistry.
Nyquist · synthetic spectrum
Bode · |Z|
Bode · phase
Choose a model, then check your reasoning.
Reference: Lazanas & Prodromidis (2023) ↗. Synthetic spectra generated from the idealized element equations; no measurement noise is included.
Review the underlying concepts
How do I know whether a fit is adequate?
Compare residual patterns (measured minus modeled, using the same sign convention) before judging a single fit score. These are schematic examples, not real fitting results.
Criterion
Meaning / units
When it is useful; warning
Residual Z′, Z″
Data − model, Ω (or normalized dimensionless)
Should be small relative to experimental noise and not show frequency-dependent structure.
Weighted sum of squares
Σ wᵢrᵢ²; units depend on weights
Compare only with the same data, weighting and residual definition.
χ² and reduced χ²
Dimensionless if residuals are divided by valid standard deviations; reduced , where N is the number of scalar residual components and p is the number of fitted parameters
Reduced χ² ≈ 1 is interpretable only with justified noise uncertainties, independent errors and appropriate model; not a universal EIS threshold.
Pseudo-χ²
Software-specific, often dimensionless
Check the software formula and weighting. A small value alone cannot certify a fit.
Parameter standard error / confidence interval
Same units as the parameter
Wide intervals mean the parameter is poorly determined; local covariance can be misleading.
Parameter correlation
Dimensionless, −1 to +1
Values near ±1 can indicate compensating parameters, not independent information.
AIC / BIC
Dimensionless model-comparison scores. For maximum likelihood L, AIC = 2p − 2 ln L and BIC = p ln N − 2 ln L. Count all estimated likelihood parameters.
Compare candidate models fitted to the same data under suitable likelihood assumptions; smaller is preferred, not proof of chemistry.
Kramers–Kronig consistency
Residual / test statistic depends on implementation
Checks compatibility with assumptions such as linearity, causality and stability; does not identify the correct circuit.
No universal acceptance number: inspect spectra and residuals, verify noise assumptions, report parameter uncertainty, test reasonable alternatives, and justify any mechanistic names independently.
Hands-on model mismatch: fit one process to two
The synthetic measurement comes from Rₛ + (R₁ ∥ C₁) + (R₂ ∥ C₂). Try to reproduce it using the deliberately inadequate Rₛ + (R₁ ∥ C₁) candidate. Both curves and complex residuals update as you adjust the candidate. A visual match is not sufficient: inspect the residuals across frequency.
Nyquist · −Im(Z) versus Re(Z) (Ω)
Bode magnitude · |Z| (Ω) versus frequency (Hz)
Real residual · Re(data − candidate) (Ω)
Imaginary residual · Im(data − candidate) (Ω)
What does the automatic search actually do?
It minimizes the unweighted sum of squared real and imaginary impedance residuals over the selected synthetic frequency window. This is a teaching approximation, not DearEIS or a statistical inference. It does not estimate uncertainty, select a physically justified model, or prove that a global optimum has been reached. Compare wide and restricted windows and inspect the residual shapes after optimizing.
Blue = synthetic measurement (Rₛ=100 Ω, R₁=1500 Ω, C₁=20 µF, R₂=800 Ω, C₂=1 µF). Orange = candidate. The synthetic data have no added noise; the residuals therefore reflect model mismatch and parameter choices, not experimental uncertainty. The exercise is manual exploration, not a numerical optimizer or a DearEIS fit.
Start with what the graph looks like
Choose a sketch first. These are idealized visual clues, not unique fingerprints of a physical mechanism.
Select a sketch to see which candidate models to test and why.
This is a teaching decision aid, not an automated physical diagnosis. A circuit is a candidate only until data quality, residuals, identifiability and physics are checked.
What electrode are you measuring?
From a spectral feature to a defensible circuit
Use this reasoning bridge after inspecting the frequency-transition analyzer. A −45° crossing is not a circuit-identification algorithm. Choose the overall spectral feature, then consider the simplest candidate and the evidence needed to test it.
Always inspect full Nyquist and Bode plots, instrument limits, repeats, and residuals. A low-frequency tail alone does not prove diffusion, and a fitted parallel resistance does not automatically establish Rct.
Software workflow: install and use
Start with PSTrace to measure and export data, then use DearEIS for quality checks and fitting. AutoEIS is an optional advanced circuit-discovery tool, not a replacement for physical interpretation.
For measurements in 1× PBS without an added redox probe, do not automatically label a fitted parallel resistance Rct. First establish whether a Faradaic reaction is actually supported. Electrode/reference geometry also affects uncompensated solution resistance.
Fit-quality statistics: definitions and limitations
Criterion
Meaning
When is it adequate?
Residuals
Measured minus predicted complex impedance; normalization and sign must be stated.
Small relative to justified noise expectations; no persistent trends, arcs, or frequency structure.
χ²
Sum of squared weighted discrepancies.
Only interpretable with the weighting and error model; raw values lack a universal pass threshold.
Reduced χ²
χ² divided by degrees of freedom, under the chosen residual definition.
Near 1 can be expected when uncertainties are correctly specified and assumptions hold; not a universal rule for EIS software.
Pseudo-χ²
Software-defined normalized misfit, often without known experimental variances.
Compare only with the same definition, weighting, dataset, and fitting setup; no universal cutoff.
Parameter uncertainty
How tightly each parameter is constrained by the data and model.
Uncertainty is small enough for the intended inference; check boundaries and unstable estimates.
Parameter correlation
Whether parameter effects can compensate for each other.
No severe confounding that makes the claimed parameter interpretation unreliable.
AIC / BIC
Relative fit-versus-complexity scores under statistical assumptions.
Useful for comparable candidates on the same data, not proof of physical mechanism.
Kramers–Kronig consistency
Checks consistency with conditions such as linearity, stability, and causality.
A useful data-quality check; passing does not validate a unique equivalent circuit.
Physical plausibility
Parameter signs, scales, mechanism, and independent evidence.
All proposed physical assignments are defensible; fit quality alone is insufficient.
Apply the reasoning
Visual fit diagnostics · recognize a misleading fit
These are synthetic teaching data, not experimental results. Change the diagnostic scenario to see why a small overall error is not enough. A residual is measured minus predicted impedance; inspect real and imaginary components separately across log frequency.
Residuals are plotted as a percentage of the synthetic observed |Z| at each frequency; this is a visualization convention, not a universal fitting objective. Frequencies span 0.1 Hz to 100 kHz.
Real-component residual
Imaginary-component residual
How to interpret: Residuals scattered around zero without structure support (but do not prove) model adequacy. Smooth sign-changing bands suggest missing physics or an unsuitable topology; isolated spikes call for checking the corresponding measurements; drift and frequency-dependent noise require experimental or weighting checks. Do not use a universal numerical residual threshold: acceptable errors depend on measurement uncertainty, weighting, frequency coverage, and the intended inference.
Residual interpretation · make the next decision
Inspect the synthetic real and imaginary residuals. Choose the most defensible next step before revealing the explanation. A residual pattern is a diagnostic clue, not proof of a mechanism.
Synthetic residuals, dimensionless; frequency axis logarithmic.
Educational synthetic examples. In real EIS fitting, examine residuals alongside measurement repeatability, weighting, stationarity, model complexity and parameter uncertainty.
Follow this checklist from a PSTrace measurement to a defensible DearEIS fit. This app does not perform the fit; it teaches how to use established software and how to judge the results.
Choose your sample
For PBS without a deliberately added redox probe, avoid assuming that a fitted resistance is charge-transfer resistance.
Publication handoff checklist
Sample construction, exposed geometric area, electrolyte composition, temperature and reference/counter electrode arrangement.
Instrument/software, AC amplitude, DC bias or OCP, frequency range, points per decade, repeats and stabilization conditions.
Original spectrum and sign convention; Nyquist and Bode plots with axes, units and measured frequency window.
Candidate circuit diagram, rationale, fitted values with units, bounds, weighting and excluded frequencies.
Real and imaginary residuals, fit criteria, uncertainty/correlation checks, and comparison with plausible alternatives.
Clearly labeled derived quantities with equations, geometry, assumptions and independent measurements; state unresolved mechanisms.
Worked reporting example · synthetic ideal Rs + C
For Rs = 1.00 kΩ and C = 1.00 µF, f₀ = 1/(2πRsC) = 159 Hz. At 1.00 kHz, |Z| = √[Rs² + (1/(2πfC))²] = 1.013 kΩ and phase = −9.0°. Report this as a synthetic ideal circuit result, not as a material property or a measured neural-electrode cutoff.
Standardized EIS reporting · Boehler et al. (2020)
Record the minimum context needed to compare spectra. The paper recommends electrode area, fcutoff, and |Z| at 1 kHz, or a full Bode plot (1 Hz–100 kHz) for recording electrodes. This is a reporting guide, not a mandate to change your validated acquisition protocol.
Report the reference electrode and counter electrode, geometry, temperature, points per decade, instrument, repetitions, and measurement date in your laboratory record as well. Do not infer fcutoff from a single 1 kHz value.
Characteristic frequency of an ideal RC arc
Start with a parallel branch R₁ ∥ C₁. Its relaxation time is τ = R₁C₁ and its characteristic frequency is f₀ = 1/(2πτ). At f₀ the isolated branch has phase −45°, magnitude R₁/√2 (−3 dB relative to its low-frequency branch value), and the Nyquist arc reaches its apex.
Adding series resistance Rₛ shifts the Nyquist arc right and changes the total impedance phase: at f₀ the total phase is −atan[(R₁/2)/(Rₛ+R₁/2)], generally not −45°. The branch's f₀ remains unchanged. A system-level −3 dB cutoff depends on the complete transfer path and is a different definition.
Nyquist · apex at f₀
Bode phase and magnitude
Definitions and interpretation limits
The ideal branch has Zₚ = R₁/(1+jωR₁C₁), |Zₚ| = R₁/√(1+(ωτ)²), and phase −atan(ωτ). The apex coordinates for the full series circuit are (Rₛ+R₁/2, −Z″=R₁/2). These exact landmarks assume one resolved ideal R∥C branch. A −45° crossing in an arbitrary measured spectrum is not a universal cutoff and does not count physical processes.
For real measurements, inspect the full magnitude and phase range. Report missing, ambiguous, or out-of-window landmarks rather than extrapolating them as measured cutoffs.
Phase transitions and −45° crossings
Explore three synthetic, illustrative spectra. The analyzer detects every phase crossing within the sampled frequency window, rather than selecting an arbitrary “cutoff.” Crossings are descriptive landmarks; neither their number nor position uniquely identifies a physical process.
Bode phase · −45° landmarks
Bode magnitude · same spectrum
Crossing
Frequency (Hz)
|Z| (Ω)
Direction as frequency increases
Crossings are located by interpolation in log frequency between sampled phase values; impedance magnitudes are interpolated in log space. A crossing between widely spaced points is only an estimate. Real experimental data require quality checks, and noisy spectra may create spurious crossings.
What additional frequencies reveal beyond 1 kHz
These two synthetic series-RC electrodes have exactly the same |Z| at 1 kHz, but different capacitance and access resistance. Their low-frequency magnitudes diverge.
Step-by-step checklist
Check items as you complete them. Progress is saved in this browser only when storage is available.
Download a clearly labeled synthetic impedance spectrum in CSV format to practice importing and plotting. The file includes frequency, Z′, Z″, −Z″, |Z|, and phase. It is not an experimental result and may need column mapping in DearEIS.
Offline acceptance package: The downloadable ZIP includes examples/EIS_Lab_synthetic_3column_signed_Zimag.csv, a six-column reference export, and DEAREIS_ACCEPTANCE.md. Use these files to perform the still-pending live DearEIS check without first navigating the simulator. The CSVs contain synthetic data, not measured electrode results.
Independent model-discrimination check (v2.8): On the bundled 61-point synthetic dataset, independently optimized Rs + (Rp ∥ C) has normalized component RMS residual 0.00111, versus 0.50061 for the inadequate Rs + C candidate (approximately 450× larger). This verifies the teaching example, not unique physical interpretation or compatibility with DearEIS. Reproduce with python tests/verify-model-discrimination.py.
DearEIS handoff · acceptance checklist
Use this checklist with a real DearEIS installation. The three-column CSV has been independently exported and numerically fitted, but its DearEIS GUI import has not been verified. Do not mark the workflow validated solely because the example looks correct in EIS Lab.
Before fitting
After fitting
0 of 6 checks completed locally. This checklist is a manual aid, not proof of validation.
Expected generating values: Rs = 100 Ω, Rp = 2000 Ω, C = 10 µF. The synthetic data include small deterministic perturbations; fitted values may differ slightly. DearEIS circuit syntax and import options depend on the installed release: follow its official documentation rather than guessing menu labels.
Guided fitting practice · from synthetic data to model assessment
This exercise mimics the reasoning after PSTrace export and before/during DearEIS fitting. It does not fit parameters automatically. Adjust a candidate model against a fixed, reproducible synthetic dataset, then inspect the residuals. Download the dataset for independent practice in DearEIS.
Import practice: The 3-column CSV contains frequency (Hz), Z′ (Ω), and signed Z″ (Ω), with no comment rows. It is a generic exchange format, not a verified DearEIS-specific format; check the importer’s column mapping and sign convention. The full CSV adds −Z″, |Z| and phase for independent inspection. The fixed synthetic data contain small deterministic perturbations; expected generating parameters are Rs = 100 Ω, Rp = 2000 Ω, C = 10 µF. A fit need not recover these values exactly.
Nyquist · fixed data versus candidate
Signed component residuals versus frequency
Practice sequence: (1) Export the synthetic CSV. (2) Confirm that its imaginary column is Z″, not −Z″. (3) Inspect the spectrum in DearEIS. (4) Compare Rs + C with Rs + (Rp ∥ C). (5) Fit the latter independently, examine residuals and parameter uncertainty, and document settings. The data contain deterministic small perturbations, so the generating values are not guaranteed to be the numerical optimum.
Residual definition: 100 × (Zdata,component − Zcandidate,component) / |Zdata|. The graph auto-scales to display all residuals. An improved visual fit does not establish a unique physical mechanism or parameter confidence.
Parameter identifiability · what the frequency window can actually determine
Explore why a visually plausible circuit can still have poorly determined parameters. This exercise uses an ideal Rs + (Rp ∥ C) circuit, not experimental data. Narrow the frequency window and watch which parameters lose sensitivity.
Normalized parameter sensitivity vs frequency
Method: sensitivities are the magnitudes of ∂Z/∂ln(parameter), divided by |Z|. The numerical condition indicator is the ratio of the largest to smallest singular value of the stacked, normalized complex sensitivity matrix. This is a local, idealized diagnostic; it is not a confidence interval, a DearEIS fit, or proof of a physical mechanism. Measurement noise, weighting, parameter bounds, model mismatch and repeated measurements also matter.
Model non-uniqueness · different circuits, similar data
Compare a reference spectrum from Rs + (Rp ∥ C) with a competing Rs + (Rp ∥ CPE) description. The CPE model contains an additional exponent n and reduces exactly to the capacitor model at n = 1. A more flexible model can fit at least as well without establishing a new physical mechanism.
Nyquist · reference and alternative
Residuals · alternative minus reference
Method: Rs = 100 Ω, Rp = 2000 Ω, C = 10 µF. For the alternative, Q = C × (2π × 100 Hz)1−n so that the capacitor and CPE admittances match in magnitude at 100 Hz. Curves are computed analytically, without noise or parameter optimization. Residuals are 100 × (Zalternative − Zreference)/|Zreference| for each component. This is a model comparison illustration, not evidence that either model represents a particular electrode mechanism.
Visual guide · good fit versus bad fit
Compare the same synthetic measured spectrum fitted with an adequate model (left) and an inadequate model (right). The dots are simulated data; the solid lines are candidate fits. Both cases use the same frequency range and plot scales.
✓ Good descriptive fit
Look for: data points follow the curve across the entire arc; residuals scatter near zero without a coherent trend.
✕ Bad descriptive fit
Look for: visible mismatches near the arc apex and edges; residuals curve systematically above and below zero.
Important: These are intentionally simplified illustrations, not DearEIS screenshots or experimental measurements. A visually good fit can still have non-identifiable parameters or the wrong physical interpretation. Compare real and imaginary residuals, data quality, and alternative circuits.
Interactive fit diagnostic microscope
Change the candidate fit while the synthetic measured points stay fixed. Compare the Nyquist overlay and the signed, normalized real and imaginary residuals. This is a teaching simulator, not a fitting algorithm.
Generating model: Rs = 50 Ω, Rp = 500 Ω, C = 10 µF. Measured points include reproducible small synthetic perturbations. Rp is a generic parallel resistance, not automatically charge-transfer resistance.
Nyquist · data vs candidate fit
Residuals vs frequency
Residual definition: 100 × (Zdata − Zfit) / |Zdata| for each component, in %. Shared axes remain fixed as you move sliders. The residual display range is ±30%; values outside it are clamped to the plot edge and flagged. This exercise illustrates systematic error, not statistical confidence or goodness-of-fit acceptance thresholds.
How to evaluate a fit
Good signs
Small, structureless residuals; physically reasonable parameters; stable results from different initial guesses; no unnecessary elements; acceptable data consistency.
Warning signs
Systematic residual arcs; parameters stuck at bounds; huge confidence intervals; multiple models fitting equally well; low-frequency drift; interpreting every fitted resistance as Rct.
Electrode case studies · connect the model to your experiment
Explore hypothetical changes between electrode states. These simulations are not measured Au, hydrogel or PEDOT spectra and do not predict a unique effect of adding a coating. Each case separates what is observed, what a model describes, and what requires independent evidence.
General theory and model interpretation: Lazanas & Prodromidis (2023) [1] ↗. Case values are pedagogical assumptions, not literature-derived material parameters.
This module teaches you to judge a fit rather than simply accept its reported error. All numbers and residuals here are synthetic teaching examples, not outputs from DearEIS or your electrodes.
Step 1 · Compare two hypothetical fitting outcomes
Real impedance residual · (data − fit) / |Z|
Imaginary impedance residual · (data − fit) / |Z|
Residuals are shown as percentages of measured |Z|. The zero line represents perfect agreement. The illustrated examples use normalized complex residuals, not a specific software's default residual definition.
Step 2 · Inspect fitted parameters
Parameter
Fitted value
Illustrative uncertainty
Interpretation
Parameter uncertainty depends on the model, weighting, noise assumptions, and fitting method. It is not a guarantee that the physical model is correct.
Step 3 · Make a scientific decision
Which next action is best supported by the evidence?
Select an action to see why it is or is not justified.
From PSTrace to DearEIS · detailed learning route
Acquire in PSTrace: record 0.1 Hz–100 kHz, applied DC potential versus Ag/AgCl or OCP, AC amplitude, electrode geometry, and electrolyte conditions. Do not assume every spectrum is stationary or linear.
Export: preserve the original data file. Identify frequency (Hz), real impedance Z′ (Ω), and imaginary impedance Z″ (Ω); check whether the export instead supplies −Z″.
Import into DearEIS: follow its official import guide and explicitly verify units and sign. Capacitive data normally appear above the horizontal axis when plotting −Z″.
Assess quality: inspect Nyquist and both Bode plots, repeat measurements, and available Kramers–Kronig consistency tools. A consistency check is informative, not a substitute for stable experimental conditions.
Fit a justified candidate: use reasonable initial guesses and bounds; document the circuit, weighting, fixed parameters and frequency exclusions. Compare simpler alternatives before adding elements.
Evaluate: inspect real and imaginary residuals across frequency, parameter uncertainty, parameter correlation, and whether multiple plausible models fit similarly.
Report: save the original spectrum, fitting settings, parameter estimates and units, residuals, figures, limitations and a description of independent experiments needed to test the mechanism.
Two mathematically different circuits can produce very similar spectra over a finite measurement window. Compare their predictions against the same synthetic measurements and decide what the data justify.
One spectrum, two candidate descriptions
Model A · one distributed relaxation
Rs + (Rp ∥ CPE)
Compact phenomenological description using a constant-phase element. It does not uniquely establish roughness, diffusion, or any particular microscopic process.
Model B · two overlapping relaxations
Rs + (R1 ∥ C1) + (R2 ∥ C2)
Two ideal RC time constants. A mathematical decomposition is not proof of two physically separate interfaces.
Nyquist · same data, two models
Real and imaginary residuals
All values are illustrative. Frequency range: 0.1 Hz to 100 kHz. Residuals = 100 × (Zdata − Zmodel) / |Zdata|. The models were selected offline to resemble each other; no fitting or parameter optimization occurs in your browser.
Make the scientific decision
The curves look similar. What can you reasonably conclude?
Select an answer to reveal the interpretation.
How would you distinguish the models?
Assess frequency-dependent residual structure, parameter confidence and correlation, repeated fits from different starting guesses, and whether an additional element materially improves predictive performance. Repeat measurements over a wider reliable frequency range, vary potential or electrolyte conditions, and test whether proposed processes change as physically expected.
Special caution for PEDOT/hydrogel electrodes: Neither model by itself establishes ion diffusion, charge transfer, polymer volumetric charging, or coating-layer contributions. A physically motivated distributed transport model may be needed, but added complexity should be justified by data.
This section tests the impedance calculator against independently written analytical identities. It checks equations and sign conventions, not whether an equivalent circuit uniquely represents an electrode. Tests run entirely in your browser.
Independent verification · documented suite
The standalone numerical verification suite checks the impedance engine against separately written analytical equations at seven frequencies (0.1 Hz to 100 kHz), including RC/CPE equivalence, finite Warburg limits, and the ideal series-RC −45° crossing. Select Run checks for the current standalone engine results. Development checks and their scope are documented in the release QA reports.
Scope: These tests validate selected circuit equations and limits, not the full interface, data-import process, real electrode physics, or DearEIS menus. Use the release QA report for current browser and software workflow results.
Analytical test suite
Run the checks to see the results.
Test
Expected behavior
Result
What passing these tests means
Checked: algebra, expected limits, phase signs, and selected analytical benchmarks for the implemented models.
Not checked: measurement accuracy, Kramers–Kronig consistency of experimental data, uniqueness of a fitted circuit, PEDOT mechanism, and correct experimental electrode geometry.
Do not use this validation panel as evidence that a circuit parameter is physically identifiable. Use the guided analysis SOP and established fitting tools for experimental analysis.
Created and developed by Ahmed Omara. EIS Lab is an independent educational application. Scientific claims and methods are referenced to the cited literature and software documentation; this acknowledgment does not imply endorsement by those sources.
[1] Lazanas, A. C. & Prodromidis, M. I. (2023). Electrochemical Impedance Spectroscopy—A Tutorial. ACS Measurement Science Au. DOI: 10.1021/acsmeasuresciau.2c00070. Covers impedance definitions, circuits, Nyquist/Bode plots, CPE and diffusion. Consult the 2025 correction when using its equations.
[2] Barsoukov, E. & Macdonald, J. R. (eds.). Impedance Spectroscopy: Theory, Experiment, and Applications. Wiley. Publisher DOI. General theory and modeling background.
[3] DearEIS documentation. Official documentation. Fitting, validation and DRT workflows.
[4] AutoEIS project. Official repository. Automated circuit discovery and ranking.
[6] Boehler, C. et al. (2020). Tutorial: guidelines for standardized performance tests for electrodes intended for neural interfaces and bioelectronics. Nature Protocols. DOI: 10.1038/s41596-020-0389-2. Box 1: recording-electrode EIS reporting; discussion of cutoff frequency and the limitations of single-frequency impedance.
Simulations implement idealized impedance equations, not measured Au, Pt, Pt/Ir, hydrogel or PEDOT properties. A CPE coefficient Q is not generally a capacitance in farads when n ≠ 1. A visually good fit does not uniquely identify physical processes.
All plots use Z = Z′ + jZ″, with capacitive impedance giving negative Z″. Nyquist plots show −Z″. The plotted Rs + (Rp || CPE) is a generic relaxation/leakage model, not proof of charge-transfer resistance.
Additional note: The frequency explorer uses illustrative bands rather than asserting a universal mapping between frequency and mechanism. A 1 kHz impedance benchmark does not mean that action potentials have a single 1 kHz frequency.
Development direction: guided PSTrace → DearEIS workflow, independently validated circuit simulations, and citations attached to electrode-specific interpretations. No fitting engine is planned for this educational application.
Compare two currents
Alternating currentDirect current
Read voltage and current together
Voltage · left axisCurrent · right axis
Follow one frequency through all three plots
This synthetic spectrum has one arc. The blue dot identifies the same sample in all three views.